UPDATES IN CARDIAC OUTPUT MONITORING

 

By

 

Saeed Mohammed Fayed, Ismail Mohammed Abdel Gawad

and Mohamed Mahmoud Mahmoud Sharaky

 

Department Of Anesthesia And Intensive Care, Faculty of Medicine, Al-Azhar University

 

Overview of cardiac output monitoring techniques (Mathews and Singh, 2008)

Invasive methods

 

-Fick's Principle

-Thermodilution technique

Minimally invasive methods

-Doppler Ultrasound

-Echocardiography

-Partial CO2 Rebreathing technique

-Indicator dilution method

-Pulse contour analysis

Noninvasive methods

-Thoracic Electrical

 Bioimpedance

-Thoracic Bioreactance

 

 

Invasive Methods                     

Fick's Principle: The first method to find cardiac output in humans was described by Adolph Fick in 1870. He postulated that oxygen uptake in the lungs is completely transferred to the blood and the total uptake or release of oxygen by lungs is the product of blood flow through the lungs and the arterio-venous oxygen content difference. Therefore cardiac output can be computed by relating oxygen consumption to arterial and mixed venous oxygen content using the equation:

 

VO2

 

(CaO2 – CvO2)

 
Cardiac Output =

Where VO2 is the oxygen content difference between inspired and exhaled gas, CaO2 is oxygen content of arterial blood and CvO2 is oxygen content of mixed venous blood (Mathews and Singh, 2008).

Pulmonary Artery Catheter (PAC): It  is a multi-lumen plastic catheter, 110 to 150 cm long with a balloon located just proximal to the tip of the distal lumen, a thermistor which measures temperature changes for the assessment of cardiac output, is located proximal to the balloon. Two additional lumens (right ventricular port and venous infusion port, if present) located at 19 cm and 30 cm from the tip reach the right ventricle and right atrium or superior vena cava. The PAC is connected to the monitoring equipment through a semi-rigid, noncompliant fluid-filled tube and pressure transducer. The movements in the transducer membrane generate electrical impulses that are amplified and transmitted to a monitor (Balk et al., 2008).

     Apart from its pressure monitoring capabilities, undoubtedly the most important feature of PAC is its ability to measure CO using the thermodilution method. The thermodilution technique has become the clinical standard for CO determination (Mathews and Singh,  2008)

Monitoring of cardiac output using PAC:

Intermittent thermodilution technique: This technique relies on principle similar to indicator dilution, but uses heat instead of color as an indicator. To do this, cold saline is rapidly injected through the proximal central venous port. A thermistor located at the distal end of the pulmonary artery catheter senses the change in temperature. Because blood flow is the source of dilution of temperature, the flow, or cardiac output, can be calculated. It is the right ventricular cardiac output that is actually measured by this technique, whereas left ventricular cardiac output can only be estimated based on the results (Pinsky, 2007)

Continuous mixed venous oximetry: The physiologic relations described by the Fick's equation form the basis for another PAC-based monitoring technique termed continuous mixed venous oximetry

 

VO2

 

 Q . 1.36 . Hgb

 
SvO2   =  SaO2    -     

     To the extent that arterial oxygen saturation, oxygen consumption, and hemoglobin concentration remain stable, mixed venous oxygen saturation may be used as an indirect indicator of cardiac output(Pinsky, 2007)

Errors in PAC: Signal loss/dampening can result from intracatheter air bubbles, debris, loose connections, vessel wall impingement, and transducer malfunction.Potential sources of error include timing, duration, volume and temperature of the injectate used. Malposition of the catheter (i.e. lodging in the RV) or displaced injectate port into the RA and misconnection of pressure monitors should be considered (Evans et al., 2009).

Complications Of PAC: Catheter site infection, bacteremia/sepsis, and right-sided infective endocarditis are well described complications in patients with indwelling venous catheters and devices, including the PAC (McGee and Gould, 2003).

Complications of CV access include arterial puncture, arterial or venous hematoma, arteriovenous fistula (AVF), pseudo aneurysm formation, thoracic duct injury, pneumothorax/hemothorax, throm-bosis and air embolization The thoracic duct may be injured during attempts to cannulate the left internal jugular or subclavian vein, potentially resulting in chylothorax or lymphatic cutaneous fistula. Attempted venous cannulation can cause pneumothorax in ~0.5% of CV catheter placement cases (Evans et al., 2009).

Contraindications of PAC: Absolute contraindications for PAC are not highlighted. However, in conditions such as cyanotic heart disease, tricuspid or pulmonary valve stenosis and severe arrythmias in which the risks outweigh benefits or measurements obtained will be redeemed inaccurate (not useful) .PAC can be considered as contraindications (Summerhill and Baram, 2005).

MINIMALLY INVASIVE METHODS

Doppler ultrasound: Ultrasound easily penetrates skin and other body tissues. As it encounters tissues of different acoustic density, a fraction of emitted ultrasound signal is reflected. When an ultrasound beam is directed along the path of the flow of blood in the aorta, using a probe, a fraction of the ultrasound signal is reflected by the moving red blood cells. The shift in the frequency of the reflected waves (Doppler shift) is proportional to the velocity of blood flow and is expressed by the equation:

Fd = 2f0/C V Cosθ

Where Fd is the change in frequency or Doppler shift, f0 is the transmitted frequency, V is the velocity of moving blood, C is the velocity of ultrasound in blood, cosθ is cosine of the angle between the direction of ultrasound beam and blood flow. The esophageal Doppler monitoring (ODM) is a widely applied method of minimally invasive cardiac output monitoring performed currently in critically ill patients  (Mathews and Singh, 2008).

    The first description of an esophageal probe measuring aortic blood flow velocity was in 1971. Suprasternal or transthoracic probes were used initially, but the difficulty of probe positioning and instability on the chest wall made their use for repeated measurements limited. This led to the development of the esophageal doppler probe. It had several advantages, including the ability to remain in position for days to weeks and its proximity to the aorta. The esophageal doppler probe is approximately the size of a nasogastric tube and can be positioned easily (Funk et al., 2009).

    In suprasternal technique, the blood flow velocity is measured in the distal aortic arch with an ultrasound transducer applied to the suprasternal notch. Because they provided only intermittent measure-ments at best, clinical acceptance of this method was limited (Thom et al., 2009).

    In transesophageal technique, a Doppler probe is inserted into the distal esophagus, and is directed to measure the blood flow in the descending aorta at about 35 to 40cm from the incisors of an intubated patient. Measurement of the Doppler frequency shift of the reflected ultrasound waves allows calculation of blood velocity. Cardiac output may then be calculated by one of two methods: The first involves measurement of the aortic cross-sectional area, measured using M mode ultrasound visualisation of the aorta and then multiplying this value by blood velocity to calculate flow. The spectral analysis of the Doppler shift gives velocity-time waveforms. A simpler, and seemingly equally reliable method is simply to derive a value of total cardiac output from a nomogram using aortic blood velocity, height, weight and age (Cholley and Singer, 2003).                                              

Transesophageal echocardiography: With this technique, cardiac output measurement is the result of calculating stroke volume which can be multiplied by heart rate. In order to assess stroke volume, it is necessary to measure flow velocity and determine the cross-sectional area. Blood leaves the left ventricular outflow tract (LVOT) as a cylinder. In order to calculate the volume of the cylinder, which is equal to the
stroke volume, the echocardiographer must obtain the diameter of the LVOT and velocity time integral (VTI) of the blood measured at the same exact location (Prahbu, 2007).

     In addition to cardiac output assese-ment, there are other uses of transeso-phageal echocardiography including managing the response to fluid resuscitation in critically ill patients at risk for heart failure or tissue hypoperfusion, critical care monitoring, pulmonary embolism therapy monitoring, prosthetic valve thrombosis monitoring, detection of myocardial ischemia , assessment of RV systolic function, and assessment of IVC size ( Porter et al., 2015).

     Ultrasonic cardiac output monitors (USCOM, Sydney, Australia) is a portable device which is non-invasive, and uses a probe placed suprasternally to measure flow through the aorta or on the left chest to measure transpulmonary flow. It uses the Doppler principle as used with ED and TEE. Main advantage is the portability of the device, and it can be used with ease in ER, OR, ICU and even in wards. Since it is a non-invasive device, it can be used by trained nursing staff and is an important screening tool for postoperative cardiac surgical patients as well (Meyer et al., 2008).

Pulse contour analysis: Pulse wave analysis is based on the principle that SV can be derived from continuous pressure waveform measurement via an arterial line. The characteristics of the arterial pressure waveform are determined by the interaction between SV and vascular compliance, aortic impedance, as well as peripheral arterial resistance (Hofer et al., 2009).

PICCO System:  incorporates a trans-pulmonary thermodilution technique (TPTD) and continuous pulse contour analysis. It is a minimally invasive technique, which gives beat-by-beat monitoring of cardiac output, and can provide accurate information on volume status and pulmonary edema (Grinberg, 2010).

     The draw back to this system is the need for a central line, recalibration every 8 hours and, cannulation of a large artery. However, it has been shown that several patient populations may benefit from central and arterial catheter measurements (Porhomayon et al., 2012).

Flo-Trac: It is another pulse contour CO monitoring system (Vigileo, Edwards Life Sciences) that was introduced in 2005. A special blood flow sensor, which is connected to an arterial line (radial, brachial, axillary or femoral artery), is needed. No external calibration is necessary (Scheeren and Wiesenack,  2008).

     Several studies have been performed concerning the accuracy of Vigileo CO monitoring that includes a variety of patients with different software versions of the device. Newer studies demonstrate a clinically acceptable precision in comparison with a standard technique of known accuracy. However, the validity depends clinically on the software version ( Yeo et al.,  2010).          

The Nexfin (BMEYE, Amsterdam, The Netherlands) is a newer device that has been introduced into practice. It provides beat-to-beat stroke volume and CO measurements by analysis of a non-invasive finger arterial blood pressure trace, derived continuously from an inflatable finger cuff. A study in 40 patients suggests that this method correlates reasonably well with transcardiopulmonary thermodilution in cardiac surgery patients (Broch et al., 2012).

The esCCO monitor (ECG–SPO2 estimated continuous cardiac output, Nihon Kohden®) is a new non-invasive tool for estimating cardiac output (CO). It derives CO from the pulse wave transit time (PWTT) estimated by the ECG and the plethysmographic wave (Bataille et al.,  2012).

Lithium indicator dilution: In this technique, a bolus of isotonic lithium chloride LiCl) solution (150mM) is injected via a central or peripheral vein, and the resulting arterial lithium concentration - time curve is recorded by withdrawing blood past a lithium sensor attached to an already existing arterial line. The CO is calculated from the lithium dose and the area under the concentration - time curve (Costa et al., 2008).

     It has been reported that it under-estimates cardiac output by 5% compared to intermittent thermodilution technique. This technique is contraindicated in patients on lithium therapy and atracurium therapy (Prahbu, 2007).

     The LiDCO system combines pulse contour analysis with lithium indicator dilution for continuous SV and SVV monitoring. The arterial pressure waveform is interpreted as a continuous curve describing the volume of the arterial tree in arbitrary units (standardized volume waveform). The effective value (approximately 0.7 times the original amplitude) of this volume waveform is determined using the root mean square, a mathematic principle to calculate the magnitude of a varying quantity. The root mean square value is called “nominal SV”, and is scaled to an “actual SV” using a patient-specific calibration factor (Montenij et al.,  2011).

NON-INVASIVE METHODS

Electrical impedance cardiography: Bioimpedance is a non-invasive technique involves the application of a small alternating current across the chest via topical electrodes. This current is thought to distribute primarily to blood because of its high electrical conductivity compared with muscle, fat and air. Pulsatile changes in thoracic blood volume result in changes in electrical impedance. The rate of change of impedance during systole is measured allowing a value of cardiac output to be derived. A number of studies have compared bioimpedance to alternative methods of cardiac output measurement, although the findings have proved inconsistent (De Waal et al., 2008).

     In order to improve the processing of the bioimpedance signal, i.e. in order to improve the signal-to-noise ratio, a modification of the thoracic electrical bioimpedance technology, the so-called ‘thoracic bioreactance’ technology has been developed. ‘Bioreactance’ represents the phase shift in voltage across the thorax (Marik, 2013).

 

 

 
   
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


 



REFERENCES

1. Balk E,  Raman G, , Chung M, , Cepeda S , Trikalinos T,  Chew P and Krishnamani R. (2008): Evaluation of the Evidence on Benefits and Harms of Pulmonary Artery Catheter Use in Critical Care Settings. Agency for healthcare Research and Quality.

2. Bataille B, Bertuit M, Mora M, Mazerolles M, Cocquet P, Masson B, Moussot PE, Ginot J, Silva S and Larché J. (2012):  Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth.,109:879–86.

3. Broch O, Renner J, Gruenewald M, Patrick M,  Jan S, Markus S,  Manu M and Berthold B. (2012): A comparison of the Nexfin(R) and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgery. Anaesthesia., 67(4):377-383.

4. Cholley BP and Singer M. (2003): Esophageal Doppler: non invasive cardiac output monitor. Echocardiography, a Journal of Cardiovascular US ,20-763-769

5. Costa MG, Della Rocca G and Chiarandini P. (2008): Continuous and intermittent cardiac output measurement in hyperdynamic condi-tions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med., 34:257–263.

6. De Waal E, Konings M, Kalkman C and Buhre W. (2008): Assessment of stroke volume index with three different bioimpedance algorithms: lack of agreement compared to thermodilution. Intensive Care Med., 34:735–739.

7. Evans DC, Doraiswamy VA and Prosciak MP. (2009): Complications Associated With Pulmonary Artery Catheters: A Comprehensive Clinical Review .Scandinavian Journal Of Surgery 98: 199–208.

8. Funk DJ, Moretti EW and Gan TJ. (2009): Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg., 108(3):887-897

9. Giraud R and Bendjelid K. (2016): Hemodynamic monitoring Int The ICU. Springer; 92-93

10. Grinberg A. (2010): The PiCCO monitor. Int Anesthesiol Clin., 48(1):57–85.

11. Hofer CK, Cecconi M, Marx G and Della Rocca G. (2009): Minimally invasive haemodynamic monitoring. Eur J Anaesthesiol., 26: 996-1002.

12. Marik PE. (2013): Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth., 27:121-34.

13. Mathews L and Singh KRK. (2008): Cardiac output monitoring. Annals of Cardiac Anesthesia, 11: 56–68.

14. McGee DC and Gould MK. (2003): Prevent-ing complications of central venous catheterization. N Engl J Med., 348:1123–1133.

15. Meyer S, Todd D, Wright I, Gortner L and Reynolds G. (2008): Review article: Non-invasive assessment of cardiac output with portable continuous-wave Doppler ultrasound. Emerg Med Australas., 20:201–208.

16. Montenij LJ, de Waal EE and Buhre WF. (2011): Arterial waveform analysis in anesthesia and critical care. Curr Opin Anaesthesiol., 24:651-656.

17. Pinsky MR. (2007): Hemodynamic evaluation and monitoring in the ICU. Chest; 132:2020–2029..

18. Porhomayon J , Ali E, Peter P and Nader D. (2012): Cardiac output monitoring devices: an analytic review. Intern Emerg Med., 7:163–171

19. Porter R, Sasha K. Shillcutt, Mark S. Adams, Georges Desjardins, Kathryn E, Joan J. Olson and Richard W. (2015): Troughton Guidelines for the Use of Echocardiography as a Monitor for Therapeutic Intervention in Adults: A Report from the American Society of Echocardiography, 28:40-56.

20. Prabhu M. (2007): Cardiac output measure-ment. Anaesthesia and Intensive Care Medicine, 8:63–67.

21. Scheeren TW and Wiesenack C. (2008): Performance of a minimally invasive cardiac output monitoring system (Flotrac/Vigileo). Br J Anaesth., 101(2):279-280.

22. Summerhill EM and Baram M. (2005): Principles of pulmonary artery catheterization in the critically ill. Lung,183:209-19.

23. Thom O, Taylor DM and Wolfe RE. (2009): Comparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter. Br J Anaesth., 103: 800-4.

24. Yeo LL, Chong S , Joshua PY and Jason P. (2010): Comparison of cardiac output measure-ment by arterial waveform analysis and pulmonary artery catheter in mitral stenosis. Ann Acad Med Singap., 39(8):655–657.

 


التحدیثات فی مجال رصد وقیاس کمیة الدم المندفع من القلب

 

سعید محمد فاید،اسماعیل محمد عبد الجواد و محمد محمود محمود شراقی.

 

قسم التخدیر والرعایة المرکزة-کلیة الطب-جامعة الأزهر

یمکن قیاس کمیة الدم المندفع من القلب بطریقتین إما نافذة أو غیر نافذة:

* الطرق النافذة:

1- طریقة  فیک (Fick) : کمیة الدم المتدفق من القلب = نسبة إستهلاک الأکسجین ÷ فرق نسبة ترکیز الأکسجین بین الشریان والورید.

2- قیاس نسبة الأکسجین فی الدم الوریدی المختلط :عندما تکون نسبة تشبع الدم الشریانی بالأکسجین, نسبة إستهلاک الأکسجین وترکیز الهیموجلوبین ثوابت، فإن نسبة الأکسجین فی الدم الوریدی المختلط تعتبر کمثال غیر مباشر لقیاس  کمیة الدم المتدفق من القلب.وهی تعتمد علی ترکیب قسطرة شریانیة رئویة.

3- إستخدام  الدلائل:   کمیة معینة من مادة خاصة ( دلیل ) تحقن فی الدورة الدمویة، ثم یتم قیاس نسبة تغیرها فی الترکیز أو الحرارة أو اللون. وهی تعتمد علی اما الحقن من خلال القسطرة الشریانیة الرئویة أو من خلال قسطرة وریدیة مرکزیة او طرفیة. ومن أشهر الدلائل محلول الملح البارد واللیثیوم. ومن أشهر الأجهزة التی تعتمد علی إستخدام الدلائل جهاز الLIDCO، والقسطرة الشریانیة الرئویة.

4- تغیر قوة النبض للضغط الشریانی:   یمکن قیاس کمیة الدم المتدفق من القلب بإستخدام الحاسب الآلی عن طریق تغیر منحنى الضغط بالشریان الأورطى أو شریان الرسغ، أو حتى بطریقة غیر نافذة من الإصبع السبابة بإستخدام مجس خاص. ومعظم هذه الطرق تحتاج إلى وضع قسطرة شریانیة طرفیة وقسطرة وریدیة مرکزیة. ویوجد العدید من الاجهزة التی تعتمد علی هذه الطریقة ومنها جهاز PICCO، وجهاز LIDCOالذی یجمع طریقتی إستخدام الدلائل وتغیر قوة النبض. وما تزال بعض الاجهزة الجدیدة محل الدراسة والمقارنة بالطرق التقلیدیة مثل جهاز Nexfin و Flo-Trac.

 

* الطرق الغیر نافذة:

          هی طرق جدیدة  لقیاس کمیة الدم المتدفق من القلب والتی لا تعتمد على ترکیب قسطرة الشریان الرئوی وبالتالی فهی لاتحتاج إلى مهارة خاصة وأدت إلى تقلیل نسبة المضاعفات التی قد تحدث،  وکذلک فهی تعطی نتائج إلى حد ما متطابقة مع مثیلاتها النافذة.

وعلى صعید آخر ،فإن هذه الطرق تلقی بعض الضوء على معلومات إضافیة عن الدورة الدمویة مثل کمیة الدم القادم إلى القلب وقوة انقباض عضلة البطین، وبالتالی فلهذه الطرق فوائد جمة علاوة على قیاس کمیة الدم المتدفق من القلب وهی تتمثل فی التشخیص المبکر وبالتالی محاولة تفادی أو سرعة إصلاح أی خلل فی الدورة الدمویة قبل التأثر الواضح الذی قد یحدث لخلایا الجسم نتیجة إنخفاض نسبة الأکسجین القادمة.

ومن هذه الطرق:

1- الموجات الفوق صوتیة ( الأشعة التلفزیونیة ) عن طریق المرئ أو من الخارج مابین عظمة القص والرقبة أو من خلال القصبة الهوائیة عن طریق الأنبوبة الحنجریة.              

2- المقاومة الکهربائیة للتجویف الصدری:  ومن الطرق الحدیثة لهذه الوسیلة توصیل الکابلات الخاصة بالجهاز إلى الأنبوبة الحنجریة. وهناک جهاز جدید یعمل بطریقة فیک (Fick) الغیر مباشرة فی قیاس کمیة الدم المتدفق من القلب وذلک بشهیق المریض الجزئی لزفیره، علاوة على أن هذه الطریقة غیر نافذة فهی سهلة للغایة ولا تحتاج سوى لتوصیل الجهاز بالأنبوبة الحنجریة ونتائجها مبهرة.

من الحکمة أن نستخدم کل الخیارات المتوفرة لمراقبة الناتج القلبی على أسس فردیة أفضل من أسس عامة، حیث أن الاضطرابات الدینامیة الدمویة لمرضى الحالات الحرجة قد تکون متعددة الأسباب وعلاج هؤلاء المرضى یجب أن یکون عملیة دینامیکیة متضمنة:

1- وسائل المراقبة.

2- التقییم.

3- إشراف وتحدید العلاج.

4- الحس الإکلینیکی.

5- إعادة التقییم.

 

وهذه النقاط الخمس یجب أن تؤخذ فی الإعتبار عند تقدیم الخدمة الطبیة لمرضى الحالات الحرجة، وأیضاً إرشاد الطبیب للتحول من وسیلة تداخلیة إلى وسیلة أقل تداخلاً والعکس صحیح    

REFERENCES
1. Balk E,  Raman G, , Chung M, , Cepeda S , Trikalinos T,  Chew P and Krishnamani R. (2008): Evaluation of the Evidence on Benefits and Harms of Pulmonary Artery Catheter Use in Critical Care Settings. Agency for healthcare Research and Quality.
2. Bataille B, Bertuit M, Mora M, Mazerolles M, Cocquet P, Masson B, Moussot PE, Ginot J, Silva S and Larché J. (2012):  Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth.,109:879–86.
3. Broch O, Renner J, Gruenewald M, Patrick M,  Jan S, Markus S,  Manu M and Berthold B. (2012): A comparison of the Nexfin(R) and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgery. Anaesthesia., 67(4):377-383.
4. Cholley BP and Singer M. (2003): Esophageal Doppler: non invasive cardiac output monitor. Echocardiography, a Journal of Cardiovascular US ,20-763-769
5. Costa MG, Della Rocca G and Chiarandini P. (2008): Continuous and intermittent cardiac output measurement in hyperdynamic condi-tions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med., 34:257–263.
6. De Waal E, Konings M, Kalkman C and Buhre W. (2008): Assessment of stroke volume index with three different bioimpedance algorithms: lack of agreement compared to thermodilution. Intensive Care Med., 34:735–739.
7. Evans DC, Doraiswamy VA and Prosciak MP. (2009): Complications Associated With Pulmonary Artery Catheters: A Comprehensive Clinical Review .Scandinavian Journal Of Surgery 98: 199–208.
8. Funk DJ, Moretti EW and Gan TJ. (2009): Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg., 108(3):887-897
9. Giraud R and Bendjelid K. (2016): Hemodynamic monitoring Int The ICU. Springer; 92-93
10. Grinberg A. (2010): The PiCCO monitor. Int Anesthesiol Clin., 48(1):57–85.
11. Hofer CK, Cecconi M, Marx G and Della Rocca G. (2009): Minimally invasive haemodynamic monitoring. Eur J Anaesthesiol., 26: 996-1002.
12. Marik PE. (2013): Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth., 27:121-34.
13. Mathews L and Singh KRK. (2008): Cardiac output monitoring. Annals of Cardiac Anesthesia, 11: 56–68.
14. McGee DC and Gould MK. (2003): Prevent-ing complications of central venous catheterization. N Engl J Med., 348:1123–1133.
15. Meyer S, Todd D, Wright I, Gortner L and Reynolds G. (2008): Review article: Non-invasive assessment of cardiac output with portable continuous-wave Doppler ultrasound. Emerg Med Australas., 20:201–208.
16. Montenij LJ, de Waal EE and Buhre WF. (2011): Arterial waveform analysis in anesthesia and critical care. Curr Opin Anaesthesiol., 24:651-656.
17. Pinsky MR. (2007): Hemodynamic evaluation and monitoring in the ICU. Chest; 132:2020–2029..
18. Porhomayon J , Ali E, Peter P and Nader D. (2012): Cardiac output monitoring devices: an analytic review. Intern Emerg Med., 7:163–171
19. Porter R, Sasha K. Shillcutt, Mark S. Adams, Georges Desjardins, Kathryn E, Joan J. Olson and Richard W. (2015): Troughton Guidelines for the Use of Echocardiography as a Monitor for Therapeutic Intervention in Adults: A Report from the American Society of Echocardiography, 28:40-56.
20. Prabhu M. (2007): Cardiac output measure-ment. Anaesthesia and Intensive Care Medicine, 8:63–67.
21. Scheeren TW and Wiesenack C. (2008): Performance of a minimally invasive cardiac output monitoring system (Flotrac/Vigileo). Br J Anaesth., 101(2):279-280.
22. Summerhill EM and Baram M. (2005): Principles of pulmonary artery catheterization in the critically ill. Lung,183:209-19.
23. Thom O, Taylor DM and Wolfe RE. (2009): Comparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter. Br J Anaesth., 103: 800-4.
24. Yeo LL, Chong S , Joshua PY and Jason P. (2010): Comparison of cardiac output measure-ment by arterial waveform analysis and pulmonary artery catheter in mitral stenosis. Ann Acad Med Singap., 39(8):655–657.